NEFA
NEFA

Registration Certificate No.: 京械注准20172400239


It can be used for quantitative determination of free fatty acids in human serum in vitro.

Clinically, it is mainly used for assistant diagnosis of hyperlipidemia, coronary heart disease and atherosclerosis.


Expected Use

It can be used for quantitative determination of free fatty acids in human serum in vitro.

Clinically, it is mainly used for assistant diagnosis of hyperlipidemia, coronary heart disease and atherosclerosis.

   

 【Testing Principle

游离脂肪酸(NEFA)测定试剂盒(ACS-ACOD法)


 【Product Characteristics

Supporting calibration products and quality control products, high detection accuracy.

Liquid double reagent, ready to use, easy to use.

Wide linear range, to meet the needs of clinical testing.

The reagent is stable for at least 14 days.


 【Performance Evaluation

1) linear

游离脂肪酸(NEFA)测定试剂盒(ACS-ACOD法)


2) Repeatability



nx(mg/L)SDCV(%)
Serum 1200.620.0071.13
Serum 2201.560.0090.58


3) Stability

Original package kit: (2-8) C, valid for 12 months

The calibration curve of the reagent can be stabilized for at least 14 days.


4) Reference interval

According to the literature and the company's test, the reference interval of NEFA is [0.10, 0.76] mmol/L.

Each hospital should establish its own reference range according to the actual situation in the region.


Recommend