It can be used to quantitatively detect homocysteine in human serum in vitro.
Hyperhomocysteinemia is an independent risk factor for atherosclerosis, thromboembolism and Alzheimer's disease.
【Product Characteristics】 |
Enzyme cycling method was used to amplify the reaction continuously with high sensitivity.
Supporting calibration products and quality control products, high detection accuracy.
The reagent is stable for at least 14 days.
1) linear
2) Repeatability
| n | x(mg/L) | SD | CV(%) |
Serum 1 | 20 | 9.88 | 0.356 | 3.60 |
Serum 1 | 20 | 20.11 | 0.273 | 1.36 |
3) Stability
Original package kit: (2-8) C, valid for 12 months
The calibration curve of the reagent can be stabilized for at least 14 days.
4) Reference interval
According to the literature, combined with our company's experimental verification, the reference interval of HCY is < 15 mol/L.
Each hospital should establish its own reference range according to the actual situation in the region.